MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG SAFIO S 60; PACING LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG SAFIO S 60; PACING LEAD Back to Search Results
Model Number 370946
Medical Device Problem Code Over-Sensing (1438)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
After an implantation period of approx.43 months, oversensing on the ventricular channel was reported.The lead was explanted.
 
Additional Manufacturer Narrative
Upon receipt, the lead under complaint was found cut 5.5 cm distal to the is-1 connector pin (into two segments).All fragments were received for analysis.An extraction aid was found on the distal fragment, it is reasonable to assume that the lead was cut during the surgery.The analysis showed abrasion marks in the distal end region, which is assumed to be the root cause of the reported oversensing.Based on the characteristics of this damage, it is reasonable to assume that the lead had been subject to severe mechanical stress in the implanted state.Significant lead motion in the area of the tricuspid valve and interaction with atypical tissues should be taken into consideration.Further analysis showed that the lead tip and ring electrode were found covered in fibrotic growth.Calcifications are known to cause high impedances and is thus assumed to be the root cause of the observed high stimulation impedance.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFIO S 60
Common Device Name
PACING LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key13972357
Report Number1028232-2022-01653
Device Sequence Number12437852
Product Code NVN
UDI-Device Identifier04035479113957
UDI-Public04035479113957
Combination Product (Y/N)N
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model Number370946
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Supplement Date Received by Manufacturer05/31/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date06/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-