| Model Number |
68HP-750 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported left side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Initial reporter name and address: (b)(6).A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is: "deflation.".
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: an opening on anterior, white particles in shell, crease flat, and wear abrasion.Leak test and microscopic analysis was performed which identified: a sharp opening on valve bond edge, a striated opening on posterior, and observed void >1 mm in non-textured, valve-to shell-bond area.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a sharp opening on valve bond edge assessed as an unidentified (tear) opening.A striated opening on posterior assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|