| Model Number |
1012272-30 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 99% stenosed lesion in the left anterior descending (lad) with heavy calcification.A 2.50 x 30mm trek balloon dilatation catheter (bdc) was advanced to the target lesion for dilatation; however, resistance was met with the anatomy.Then during the first inflation, the balloon ruptured at 8 atmospheres.The bdc was removed from the patient without issue.There was no adverse patient effect reported.The procedure continued and the target lesion was successfully treated with a 2.5x33mm stent.No additional information was provided.
|
| |
|