• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN® EP® HEALING ABUTMENT 5MM(D) X 5.6MM(P) X 6MM(H); DENTAL HEALING ABUTMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN® EP® HEALING ABUTMENT 5MM(D) X 5.6MM(P) X 6MM(H); DENTAL HEALING ABUTMENTS Back to Search Results
Model Number IWTH56
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during uncovering, the healing abutments did not seat.No impact to the patient reported.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).Patient identifier unknown / not provided.Age and date of birth unknown / not provided.Patient sex unknown / not provided.Weight unknown / not provided.
 
Event or Problem Description
No further event information is available at the time of the report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The following fields have been updated: b4: date of this report b5: describe event or problem d4: additional device information g3: date received by manufacturer g6: type of report h1: type of reportable event h2: follow up type h3: device evaluated by manufacturer h6: adverse event problem h10: additional narrative two healing abutments (iwth56 and imha36) were returned for investigation.Visual evaluation of the as returned products identified no malfunction/damages.Functional testing was performed using in-house mating device and no malfunction was identified.Additionally, measurements were taken using a caliper.Through dimensional analysis and comparison to drawings, the devices were determined to be within design specifications.Device history record (dhr) review for the lots (1248387 and 1249645) had revealed no deviations nor non-conformances which could have caused or contributed to the reported event.All products were conforming at the time they left zimvie.Lots were inspected and passed all acceptance criteria by qa.Complaint history review was performed for the lots (1248387 and 1249645) for similar events and no other complaint about nonconforming products was identified.Therefore, based on the available information and functional testing, device malfunction has not occurred and the reported event has been unconfirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® EP® HEALING ABUTMENT 5MM(D) X 5.6MM(P) X 6MM(H)
Common Device Name
DENTAL HEALING ABUTMENTS
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key13974136
Report Number0001038806-2022-00471
Device Sequence Number4789460
Product Code NHA
UDI-Device Identifier00844868009191
UDI-Public(01)00844868009191(17)260929(10)1248387
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K072642
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/29/2026
Device Model NumberIWTH56
Device Catalogue NumberIWTH56
Device Lot Number1248387
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer07/11/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-