Model Number 75004650
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: case-(b)(4).
 
Event or Problem Description
It was reported that, during a tha, the polarstem stem inserter broke.It is unknown if any fragment fell inside of the patient.Surgery was completed freehandedly with a delay of less than 30 min.Patient was not harmed.
 
Additional Manufacturer Narrative
H3, h6: it was reported that, during a total hip replacement, the polarstem stem inserter broke.The device, intended for use in treatment, was not returned for investigation.Therefore, a product evaluation is not possible.The production history of the device was reviewed.There are no indications that the device failed to match specification at the time of manufacturing.A complaint history review was conducted.The complaint is in line with the current risk management.Review of past corrective actions was performed.No further escalation is required.It must be noted, that this stem inserter shall only be used to implant the cemented version of the polarstem.The device should never be impacted to insert the stem.Due to insufficient information, the relationship between the reported event and the device cannot be confirmed.The root cause cannot be determined and we cannot speculate about factors which would have contributed to the reported issue.Should the part be returned for investigation, the investigation will be reopened.No further actions are deemed necessary.Smith and nephew will monitor this device for further similar issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARSTEM STEM INSERTER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13975130
Report Number9613369-2022-00072
Device Sequence Number14646721
Product Code LXH
UDI-Device Identifier07611996097573
UDI-Public07611996097573
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75004650
Device Catalogue Number75004650
Device Lot NumberD62575
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-