Catalog Number ROB100022
Medical Device Problem Codes Failure to Power Up (1476); Sparking (2595)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during set up for a cori demonstration, sparks were seen coming from the transformer underneath and to the rear of the real intelligence robotics cart base 220.Then, the real intelligence cori then failed to turn on.No case involved; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Results of investigation: the product, real intelligence robotics cart base 220, rob100022, s/n (b)(6), intended for use in treatment was not returned for evaluation, however a phone interview with the uk technical support team was conducted.A relationship between the reported event and the device was established.The reported problem was visually and functionally confirmed.Sparks were seen below the base when the cori power cord was plugged into wall power.The cori power cord was replaced and the system powers up with no issues.The most likely cause of the event is product mishandling.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints found one similar event.A historical escalation event review was completed.A review of prior escalation actions found no actions applicable to the scope of the reported complaint.Refer to the cori instructions for use for proper handling.Care and caution should be exercised when storing the power cord and moving the cori product between locations.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still adequate.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTICS CART BASE 220
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13975513
Report Number3010266064-2022-00236
Device Sequence Number6284572
Product Code OLO
UDI-Device Identifier00885556774151
UDI-Public885556774151
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberROB100022
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer04/05/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date04/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/05/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
REAL INTELLIGENCE CORI-SN (B)(6); ROB10024, REAL INTELLIGENCE CORI