Batch review performed on 25 march 2022.Lot 2106489: (b)(4) items manufactured and released on 29-jul-2020.Expiration date: 2026-07-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.Clinical evaluation performed by medical affairs director: early revision surgery, few months after primary, of a cemented tka due to difficulties in flexion.The new surgeon repositioned the femoral component, downsizing it by 1 size, and placed a slightly thicker insert, leaving the tibial component in place.This problem is not due to a device failure.Other device involved: batch review performed on 25 march 2022: gmk-sphere 02.12.0210fl tibial insert fixed sphere flex #2/10 mm l (k121416) lot 2104644: (b)(4) items manufactured and released on 08-jun-2021.Expiration date: 2026-05-24.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|