• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ACT ARTIC E1 HIP BRG 28X38MM S44 DIA28; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ACT ARTIC E1 HIP BRG 28X38MM S44 DIA28; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Concomitant medical products: catalog number: ep-200144; lot number: 926670; brand name: act artic e1 hip brg.Catalog number: 650-1055; lot number: 2020060247; brand name: cer bioloxd option hd.Catalog number: 650-1064; lot number: 2997131; brand name: cer option type 1 tpr.Catalog number: 51-102070; lot number: 6663490; brand name: tprlc xr fp type1 pps.Catalog number: 110010262; lot number: 110010262; brand name: g7 osseoti multihole 48mm.Catalog number: 110024461; lot number: 814540; brand name: g7 dual mobility liner.Multiple reports were reported along with this report 0001825034-2022-00795.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the patient was revised due to dislocation approximately 1 year post implantation.Attempts have been made and additional information on the reported event is unavailable.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were visual examination of the provided pictures identified the bearing is deformed on the outer lip and has scratching.The femoral head has black discoloration on one side.Complaint sample was evaluated and the reported event was not confirmed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that the patient was revised due to disassociation of the head and bearing approximately 1 year post implantation.Attempts have been made and additional information on the reported event is unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACT ARTIC E1 HIP BRG 28X38MM S44 DIA28
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13977642
Report Number0001825034-2022-00711
Device Sequence Number5555086
Product Code KWY
UDI-Device Identifier00880304485020
UDI-Public(01)00880304485020(17)250624(10)926670
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161190
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2025
Device Model NumberN/A
Device Catalogue NumberEP-200144
Device Lot Number926670
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer05/06/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-