During the processing of this complaint, attempts were made to obtain complete event, patient, and device information.It was reported that during the attempt to implant the lens using the yamane technique, the lens could not be secured.Thus, the lens was explanted and an anterior chamber lens was used instead.No adverse patient effects or clinically significant delay in procedure reported.No additional information provided.It was stated by the customer that they are using the yamane technique to implant lenses.This technique induces a larger amount of force while trying to position the haptics in the scleral tunnel.Our lenses are intended to be implanted in the capsular bag; therefore, the lenses are being used off-label.The device history records were reviewed and there were no non-conformities or deviations noted during the manufacturing of the lenses that would have contributed to the nature of this complaint.Additionally, lenses are 100 % inspected before they leave our manufacturing site.Therefore, we are confident that the lens was processed per standard operation procedures and inspections and met all of the criteria for release.Additionally, there was no damage noted to the lens during preparation for use which suggest a product deficiency did not contribute to the reported issue.Based on the information provided, the risk assessment for the lucia product and based on our experience and other complaints that have already been resolved we have determined that the following factor may have cause or contributed to the issue are known to be caused by numerous factors including, but not limited to: loading strategy, lens placement technique, poor handling during folding and inserting, patient pathology.We have reviewed the information provided and at this time there is no indication of a product quality issue with respect to the manufacturing, design, or labeling of the device.It appears that the reported issue is linked to off label use.However, the need to explant the lens is considered need for medical intervention in order to prevent and avoid permanent impairment or damage to a body structure.
|
|
It was reported that during the attempt to implant the lens using the yamane technique, the lens could not be secured.Thus, the lens was explanted and an anterior chamber lens was used instead.No adverse patient effects or clinically significant delay in procedure reported.No additional information provided.
|