| Model Number |
MX448HFB |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported on medwatch (b)(4) iv lipid filter not allowing lipids to pass through from iv tubing into the filter.Changed filter with different lot number, and still did not work.No additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Other text: one unit was returned for investigation.Upon physical inspection, it was found that the complained issue could be duplicated.The leakage is related to a supplier issue.A dhr review was done and no issues related to the original issue were found.
|
| |
|