| Model Number |
72007 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No additional information was made available at the time of reporting.
|
| |
|
Event or Problem Description
|
|
Patient reported experiencing pain since on (b)(6) 2021 and claims to not receiving adequate pain relief from the system.Reprogramming attempts were not able to resolve issue.Ipg appeared to have migrated and 4-contacts appeared to be broken at the tines.System was explanted on (b)(6) 2022.
|
| |
|
Search Alerts/Recalls
|