Model Number 190713
Device Problems Thermal Decomposition of Device (1071); Melted (1385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2022
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned and melted distribution board.The distribution board was burned and melted where it connects to the power supply.The biomed also reported that the heater rod was not working.The burned board was noticed during machine repair after the machine was initially pulled from service for not reaching appropriate temperature.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 17,150 hours and the distribution board, power supply, and heater rod were original to the fresenius machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted distribution board.The biomed replaced the distribution board, power supply, and heater rod, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The distribution board and heater rod were discarded by the biomed however the power supply is available to be returned to the manufacturer for physical evaluation.
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned and melted distribution board.The distribution board was burned and melted where it connects to the power supply.The biomed also reported that the heater rod was not working.The burned board was noticed during machine repair after the machine was initially pulled from service for not reaching appropriate temperature.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The machine has approximately 17,150 hours and the distribution board, power supply, and heater rod were original to the fresenius machine.The biomed reported that there was no damage observed on any other components, or any other additional issues, associated with the burned and melted distribution board.The biomed replaced the distribution board, power supply, and heater rod, which resolved the issue.The unit was returned to service at the user facility without issue and without reoccurrence of the event.The distribution board and heater rod were discarded by the biomed however the power supply is available to be returned to the manufacturer for physical evaluation.
 
Manufacturer Narrative
Additional information: d9, h3 plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius field service technician (fst).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYS., WITH CDX
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
Manufacturer Contact
jason busch
920 winter st
waltham, MA 02451
9043166958
MDR Report Key13978322
MDR Text Key288393117
Report Number2937457-2022-00559
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100897
UDI-Public00840861100897
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093902
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number190713
Device Catalogue Number190713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device AgeMO
Date Manufacturer Received04/04/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-