| Model Number |
1012632-39 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional omnilink stent delivery system (sds) (1012632-29, 1030941) is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the right common iliac artery.During advancing of the first omnilink stent delivery system (sds) (1012632-39, 1050441), the stent dislodged, and a snare device was used to retrieve the dislodged stent successfully.A second omnilink sds (1012632-29, 1030941) was advanced into the vessel and the stent dislodged as well.The dislodged stent was able to be retrieved with the delivery system but a cut down of the vessel was performed to remove the second dislodged stent.There was some resistance noted during advancement due to a sharp angle in the vessel.A third omnilink sds was used successfully to complete the procedure.There was no adverse patient sequela.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined the reported difficulties and subsequent removal of foreign body appear to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|