| Model Number |
1012632-29 |
| Medical Device Problem Codes |
Failure to Advance (2524); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional omnilink stent delivery system (sds) (1012632-39, 1050441) is filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the right common iliac artery.During advancing of the first omnilink stent delivery system (sds) (1012632-39, 1050441), the stent dislodged, and a snare device was used to retrieve the dislodged stent successfully.A second omnilink sds (1012632-29, 1030941) was advanced into the vessel and the stent dislodged as well.The dislodged stent was able to be retrieved with the delivery system but a cut down of the vessel was performed to remove the second dislodged stent.There was some resistance noted during advancement due to a sharp angle in the vessel.A third omnilink sds was used successfully to complete the procedure.There was no adverse patient sequela.No additional information was provided.
|
| |
|
Event or Problem Description
|
|
Medwatch report received states, stent came off of stent balloon in the right common iliac artery and was partially removed using balloon trapping, but was unable to be removed from the left femoral arteriotomy site requiring surgical exploration and foreign body extraction followed by direct repair of the left femoral arteriotomy site with probable perioperative blood loss anemia.Per the physician, there was no bleeding complication during the procedure and no intervention performed.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties and subsequent surgical intervention appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Attachment: user facility medwatch report.
|
| |
|