• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012632-29
Medical Device Problem Codes Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.The additional omnilink stent delivery system (sds) (1012632-39, 1050441) is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the right common iliac artery.During advancing of the first omnilink stent delivery system (sds) (1012632-39, 1050441), the stent dislodged, and a snare device was used to retrieve the dislodged stent successfully.A second omnilink sds (1012632-29, 1030941) was advanced into the vessel and the stent dislodged as well.The dislodged stent was able to be retrieved with the delivery system but a cut down of the vessel was performed to remove the second dislodged stent.There was some resistance noted during advancement due to a sharp angle in the vessel.A third omnilink sds was used successfully to complete the procedure.There was no adverse patient sequela.No additional information was provided.
 
Event or Problem Description
Medwatch report received states, stent came off of stent balloon in the right common iliac artery and was partially removed using balloon trapping, but was unable to be removed from the left femoral arteriotomy site requiring surgical exploration and foreign body extraction followed by direct repair of the left femoral arteriotomy site with probable perioperative blood loss anemia.Per the physician, there was no bleeding complication during the procedure and no intervention performed.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents.The investigation determined the reported difficulties and subsequent surgical intervention appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Attachment: user facility medwatch report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key13978528
Report Number2024168-2022-03474
Device Sequence Number5567616
Product Code NIO
UDI-Device Identifier08717648178542
UDI-Public08717648178542
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2024
Device Model Number1012632-29
Device Catalogue Number1012632-29
Device Lot Number1030941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-