The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging black particles related to a bipap device's sound abatement foam.There was no report of serious or permanent harm or injury.The device was returned to the third-party service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the third-party.There was one error found.The third-party concludes that they could not confirm the customer's allegation and there was no visible foam degradation and unit got scrapped.In this report, section g4 has been corrected.In this report, section d8, d9, h2, h3, and h6 has been updated.
|