A.I.D.D LONGFORD ARCHITECT B12 REAGENT KIT; RADIOASSAY, VITAMIN B12 Back to Search Results
Model Number 7K61-27
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed falsely elevated architect b12 results for multiple samples when compared to a reference laboratory.Additionally, the customer noted there was a shift in qc results starting on (b)(6) 2022.The following data was provided (customer¿s reference range: 211 to 1074 pg/ml, test date was (b)(6) 2022, units of measure is pg/ml): sid (b)(6) architect result = 315, arup result = 200.Sample 5: architect result = 360, arup result = 255.Sample 6: architect result = 383, arup result = 273.Sample 10: architect result = 405, arup result = 254.Sample 14: architect result = 371, arup result = 235.No impact to patient management was reported.
 
Additional Manufacturer Narrative
The complaint investigation for falsely elevated architect b12 results included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, and device history record review and in-house testing.Trending review did not identify any trends for the issue for the product.Device history record review did not identify any nonconformances or deviations associated with the likely lot number 34043ud00 and complaint issue.In-house accuracy testing was completed with complaint lot number 34043ud00.Testing met validity and acceptance criteria and the product is performing as expected.Labeling was reviewed and sufficiently addresses the customer's issue.Based on our investigation, no systemic issue or deficiency with the architect b12 reagent for lot 34043ud00 was identified.
 
Event or Problem Description
The customer observed falsely elevated architect b12 results for multiple samples when compared to a reference laboratory.Additionally, the customer noted there was a shift in qc results starting on 03mar2022.The following data was provided (customer¿s reference range: 211 to 1074 pg/ml, test date was (b)(6) 2022, units of measure is pg/ml): sid (b)(6) initial result = 315, arup result = 200.Sample 5: initial result = 360, arup result = 255.Sample 6: initial result = 383, arup result = 273.Sample 10: initial result = 405, arup result = 254.Sample 14: initial result = 371, arup result = 235.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT B12 REAGENT KIT
Common Device Name
RADIOASSAY, VITAMIN B12
Manufacturer (Section D)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI  N39E932
Manufacturer (Section G)
A.I.D.D LONGFORD
lisnamuck
co. longford
longford N39E9 32
EI   N39E932
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key13978736
Report Number3005094123-2022-00063
Device Sequence Number5440258
Product Code CDD
UDI-Device Identifier00380740014179
UDI-Public00380740014179
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K121314
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/27/2022
Device Model Number7K61-27
Device Catalogue Number07K61-27
Device Lot Number34043UD00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer06/01/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date06/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/13/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC I2000SR INST, 03M74-02, ISR62340; ARC I2000SR INST, 03M74-02, ISR62340
Patient SexUnknown
-
-