• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPRINT QUATTRO SECURE MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPRINT QUATTRO SECURE MRI SURESCAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number 694765
Device Problems Signal Artifact/Noise (1036); Fracture (1260); High impedance (1291); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems Distress (2329); Shock from Patient Lead(s) (3162)
Event Date 12/08/2021
Event Type  Injury  
Event Description
It was reported by the patient that the device was defective and had been inactivated but could not be replaced due to potential bleeding problems.It was determined that the patient had been inappropriately shocked due to noise on the right ventricular (rv) lead and has since suffered post-traumatic stress.The patient's system was inactivated.The decision was made to not replace the lead as the patient's ejection fraction had improved.The lead remains in situ. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Concomitant medical product: biotronik crt-d, implant date: unknown.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was further reported that the rv lead also had elevated impedance and there was a concern for a fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE MRI SURESCAN
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13978837
MDR Text Key288472956
Report Number2649622-2022-06191
Device Sequence Number1
Product Code LWS
UDI-Device Identifier00613994249890
UDI-Public00613994249890
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P920015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/18/2010
Device Model Number694765
Device Catalogue Number694765
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/08/2022
Initial Date FDA Received04/01/2022
Supplement Dates Manufacturer Received05/30/2023
Supplement Dates FDA Received05/31/2023
Date Device Manufactured03/28/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
419588 LEAD, 5568-45 LEAD.
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-