Model Number N/A
Medical Device Problem Codes Material Frayed (1262); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported guide wire frayed thus making it impossible to pass the catheter.No other information was provided.
 
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of redt4307 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported guide wire frayed thus making it impossible to pass the catheter.No other information was provided.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), applicable manufacturing records, photo and video analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a frayed wire was confirmed but the cause is unknown.Two photographs and a video were returned for evaluation.What appeared to be a braided stylet wire was seen in the photos and video.Review of the damaged stylet revealed that the pull tab was missing from the wire.The wire was elongated and frayed with separation between the inner core wire and the outer coil wire.The wire contained multiple kinks and bends which may indicate difficult insertion or removal.This type of damage can occur if the wire fractures or is cut.Possible contributing factors could include difficult stylet removal due to not hydrating the hydrophilic coating, sharp instrument damage to the stylet and/or removing the stylet against resistance.H3 other text : evaluation findings are in section h.11.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a broken stylet is confirmed.One 0.015 in.Stiffening stylet was returned for evaluation.An initial visual observation showed use residue on the returned sample.The stylet was observed to be bent in many locations and the core wire of the stylet was observed to be broken with the majority of the coiled wire unraveled.The proximal end of the stylet was broken off and missing.A microscopic observation revealed the fracture site of the core wire to be angled, lustrous, and peaked, and linear striations were observed leading up to the peak of the break.The break site of the coiled wire was observed to be mostly flat and smooth.The observed characteristics of the break sites in the returned stylet are indicative of damage caused by a sharp instrument such as scissors.H3 other text : evaluation findings are in section h.11.
 
Event or Problem Description
It was reported guide wire frayed thus making it impossible to pass the catheter.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PER-Q-CATH PLUS 4F SINGLE-LUMEN PICC
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13978905
Report Number3006260740-2022-01099
Device Sequence Number17650253
Product Code LJS
UDI-Device Identifier00801741074875
UDI-Public(01)00801741074875
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K954104
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2022
Device Model NumberN/A
Device Catalogue Number4134105
Device Lot NumberREDT4307
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer04/28/2022
08/12/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date05/07/2022
08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age89 YR
Patient SexFemale
-
-