Model Number 602-90780
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the knob of an aleutian an inserter fractured off from the shaft intra-operatively.Surgery was successfully completed and no adverse consequences or medical intervention were reported.
 
Event or Problem Description
It was reported that the knob of an aleutian an inserter fractured off from the shaft intra-operatively.Surgery was successfully completed and no adverse consequences or medical intervention were reported.
 
Additional Manufacturer Narrative
A visual inspection was performed on the device and the reported issue of fracture of the proximal knob was confirmed.The inner shaft was able to be taken out.Device history records were not reviewed for the corresponding lot so no relevant issues were identified.A review of complaint history associated with the subject catalog and lot number was performed, and no adverse trends were observed.It was initially reported that the top of the aleutian an inserter inner shaft sheared off during impaction.Cascadia an view and guide surgical technique was reviewed: ¿the proximal end of the implant should be aligned with the distal tip of the inserter so the fingers on the inserter line interbody insertion up with the recesses on the implant and the angled surfaces of the implant and inserter align.The implant can then be threaded onto the inserter by turning the thumbwheel at the proximal end of the inner shaft clockwise." the damage on the returned device evidenced knob fracture on the proximal end of the inserter.The most likely case for this failure mode based on the deformation observed is excessive torsional force.As the customer reported that the inserter fractured when malleating this is most likely because the inserter was at the yielding point from the torsional force applied with the mallet causing it to go over that threshold and fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AN INSERTER INNER SHAFT
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key13979860
Report Number3004774118-2022-00127
Device Sequence Number5568222
Product Code MAX
UDI-Device Identifier10888857157255
UDI-Public10888857157255
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K160547
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number602-90780
Device Catalogue Number602-90780
Device Lot NumberNHKJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer06/09/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date07/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-