(b)(4).Batch # u5ce50.Component code: (b)(4).Additional information was requested, and the following was obtained: is the current patient status known? -unknown.Was there any patient consequence or change in the post-operative care of the patient as a result of the event? (extended hospital stay, readmission, re-operation, etc.) -unknown.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one ple60a device was returned with no apparent damage and with no reload present.Upon evaluation, the device would not fire.In order to verify the condition of the internal components, the device was disassembled to verify the condition of internal components and the pcb board was noted to be non-functional.No conclusion could be reached as to what may have caused the pcb board to be damaged.As part of our quality process, the manufacturing records of this batch number were reviewed, and the manufacturing standards were met prior to the release of this batch,.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|