MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
3544325 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
11/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6)-year-old caucasian female patient underwent a breast augmentation primary with mentor memorygel breast implants 350cc (r) and 325cc (l) and reported rupture on the right side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent removal and replacement with similar devices on (b)(6) 2022.Two devices were received damaged.As a result, both devices will be conservatively reported for bilateral rupture in the absence of additional information.This report is for the left breast prosthesis.See manufacturer report number 1645337-2022-02408 for contralateral device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the gel siltex 325cc breast implant was found to have an area of delamination at the union between the patch and shell, causing gel leakage.Although potential causes of patch delamination are unknown, stresses and forces on the implant in-vivo or exposure to betadine at insertion could result in delamination and ruptures.Per the current product insert data sheet (pids), rupture is a known complication associated with these devices and can occur at any time during or after implantation.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Manufacturer¿s reference number: (b)(4).
|
| |
|