| Model Number |
1012455-08 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a 50% stenosed lesion in the right coronary artery (rca) with heavy calcification and heavy tortuosity.A 5.00 x 8mm nc trek balloon dilatation catheter (bdc) was advanced to the target lesion; however the bdc met with resistance during advancement due to anatomical challenges.During the first inflation the balloon ruptured at 14 atmospheres; therefore, bdc was removed from the patient.A non-abbott balloon was used to complete the procedure.There was no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|