• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC YUEH CENTESIS DISPOSABLE CATHETER NEEDLE; GCB NEEDLE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC YUEH CENTESIS DISPOSABLE CATHETER NEEDLE; GCB NEEDLE, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/23/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a yueh centesis disposable catheter needle separated.The device was required for a computed tomography (ct) guided drain placement into a pelvic abscess and was placed via direct stick.As the physician was removing the device, and the hub broke off.The device was then removed in its entirety, as the access wire was still in place.The procedure was completed "as normal" and patient outcome was "as expected".The customer noted there was no difficult advancement or force exerted on the device.No obstructions were encountered.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.
 
Additional Manufacturer Narrative
User facility reference number: (b)(4).Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: it was reported that a yueh centesis disposable catheter needle separated.The device was required for a computed tomography (ct) guided drain placement into a pelvic abscess and was placed via direct stick.As the physician was removing the device, the hub broke off.The device was then removed in its entirety, as the access wire was still in place.The procedure was completed "as normal" and patient outcome was "as expected." the customer noted there was no difficult advancement or force exerted on the device.No obstructions were encountered.As reported, the patient did not experience any adverse effects or require any additional procedures due to this occurrence.Reviews the complaint history, device history record (dhr), quality control procedures, and a supplier investigation were conducted during the investigation.The complainant did not return the complaint device to cook for investigation.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhrs for the reported complaint device lot 9488736 and the related subassembly lots revealed no recorded non-conformances relevant to the failure mode.Additionally, a complaint data base search did not find any additional complaints from the lot.Cook did not identify any nonconforming material in house or in the field.Cook also reviewed product labeling.This complaint device is not currently packaged with instructions for use.The affected component is supplied from merit medical systems/malvern div.Cook requested the supplier investigate this matter.The supplier reviewed the device history record for this device.The supplier did not find any deviations or nonconformances associated with this lot.The supplier confirmed that procedural and functional requirements needed for this product release were met.A failure mode could not be verified due to no device return or photos.It was reported that during the procedure, the hub of the catheter broke off in the patient.The device was able to be removed as an entirety.The procedure was completed as normal.The patient did not experience any adverse effects.Based on the returned supplier investigation, cook concluded the cause of event to be component failure.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
YUEH CENTESIS DISPOSABLE CATHETER NEEDLE
Common Device Name
GCB NEEDLE, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key13986529
Report Number1820334-2022-00523
Device Sequence Number5413149
Product Code GCB
UDI-Device Identifier00827002094895
UDI-Public(01)00827002094895(17)220131(10)9488736
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model NumberN/A
Device Catalogue NumberDTVN-5.0-19-7.0-YUEH
Device Lot Number9488736
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/02/2022
Supplement Date Received by Manufacturer09/06/2022
Initial Report FDA Received Date04/01/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/31/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
Patient Weight134 KG
Patient RaceWhite
-
-