| Model Number |
CB96670-019 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Hemorrhage/Blood Loss/Bleeding (1888); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that during use of a bio-medicus cannula, the customer observed a crack on the cannula body and a leak was observed from the cannula.The device was replaced to complete the procedure.There was no patient impact associated with this event.Additional information received states that the cannula was not heated or cooled prior to use.There was no damage in the location of the sutures.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: medtronic cannot confirm or deny the complaint of a crack in the cannula body and the cannula leaking blood as no product has been returned to date.A root cause of this occurrence cannot be determined without returned product.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Trends for issues with this device are monitored.Correction b5: medtronic received additional information stating that the amount of blood lost as a result of the leak is unknown.The customer stated that it is also unknown whether a transfusion was required because of the leak.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|