| Model Number |
ETLW1624C124E |
| Medical Device Problem Codes |
Leak/Splash (1354); Migration or Expulsion of Device (1395); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii limb stent graft was implanted during the endovascular treatment of a 55mm abdominal aortic aneurysm.It was reported on an unknown date follow up scan, it was identified that the left common iliac stent had migrated proximally losing apposition with the iliac wall resulting in a type ib endoleak.Approximately 9 months post the index procedure an etlw1624c124e was implanted into the left common iliac artery.After angiogram it was noted that the endoleak was resolved and the case was concluded.Per physician, the migration was caused by initial aortic remodeling.No additional clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
H6.Evaluation conclusion code updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|