SMITHS MEDICAL INTERNATIONAL LTD PORTEX; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
| Model Number |
370900-NL |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Disposable anesthesia breathing circuit failed compliance tests multiple times after using 3 circuits from same lot number.When manual leak test performed, circuit would only hold pressure if apl was set to 40cmh2o or greater.When circuit was released below 40cmh2o breathing circuit would not remain pressurized.
|
| |
|
Search Alerts/Recalls
|
|
|