As a result of an fda inspection conducted in jan 2020, exactech, (b)(4), has committed to remediating 3 years of complaints (2017-2020).This mdr is being submitted as part of that remediation.Information was not provided for: visual evaluation condition/findings (conducted with a 7x or l0x optical unless otherwise specified: the fractured head of the mini headed fixation pin appears consistent with applying a bending moment to the proximal head of the pin with the pin puller instrument.Tension was likely applied to the pin puller at an angle rather than axially as intended.All other aspects of the devices appear unremarkable and consistent with years of use.Design-related issues: the design of the mini headed fixation pin has been in the field since 1999.Exactech is aware of 5 complaint reports involving 6 broken mini headed fixation pins since 2008.The frequency of occurrence ranking scale is very low; therefore, this issue does not appear to be design related.Mfg-related issues: manufacturing related data could not be reviewed because no information regarding the manufacturing lots or serial numbers were provided.Corrective actions are not required because the occurrence rate is "very low", and the risk is captured in the rmr.The broken pin reported was likely the result of applying a bending moment to the head of the pin with the pin puller, which allowed for the proximal rim to fracture.Ifu 700-096-181: instrument inspection ¿ visually inspect the instruments for damage such as fractures; cracks; gouges; deformation; burrs; discoloration, corrosion, or rust; excessive component wear; nicks on cutting surfaces, missing or loose components; blockages in cannulae, cleaning holes or other cavities that cannot be removed via standard cleaning; worn or difficult to read markings/engravings; or other apparent damage.Check the function of mechanisms by actuating any levers, knobs, switches, connectors, sliding features, hinges, or other mechanical interface features.Ensure smooth operation of these features over their functional range of motion.If damage, wear, or non-functioning/poorly functioning mechanisms are found, do not use the instrument, and contact the sales representative or customer service for disposition.Ifu states: the surgeon shall become thoroughly familiar with the technique of implantation of the prostheses by: appropriate reading of the literature, and training in the operative skills and techniques required for surgery, and reviewing information regarding use of instrumentation.It is a clinical standard of practice in the operating room that all instruments should be visually and functionally inspected before use, these guidelines are from standard association of peri-operative registered nurses (aorn) guidelines.Surgeons are to be familiar and knowledgeable with all instrumentation, devices and proficient with surgical techniques.This device is used for treatment not diagnosis.Based on review of all available information, there is no evidence to reasonably suggest the reported event is related to any design issues.There was no reported patient adverse event.No device fragments were left in the patient.An investigation was conducted; the broken pin reported was likely the result of applying a bending moment to the head of the pin with the pin puller, which allowed for the proximal rim to fracture.
|