• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. BUR,OVAL,FLUSHCUT,12 FLUT,5.0MMX 13CM; BURR, SURGICAL, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. BUR,OVAL,FLUSHCUT,12 FLUT,5.0MMX 13CM; BURR, SURGICAL, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number BUR,OVAL,FLUSHCUT,12 FLUT,5.0MMX 13CM
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported on (b)(6) 2022 by a facility representative via sems that an ar-8500fot burr bent and twisted.This occurred during a case and broke in the joint space.Per facility: burr was being used in the posterior tibia.The surgeon experienced shavings from the burr.The guard on the burr bent and twisted around.The effect on the patient was metal shavers in the knee joint due to the guard bending and twisting.".
 
Additional Manufacturer Narrative
Complaint confirmed.Upon visual evaluation, no problem was noted with the (b)(4) packaged devices.The (b)(4) unpackaged devices were found to have the guard twisted and bent and on one device the burr came apart from the device.It was also noted that there was discoloration on the (b)(4) unpackaged devices.Based on the observed condition of the device, this event is most likely the result of user applied forces via prying/leveraging against bone and/or hard tissue during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BUR,OVAL,FLUSHCUT,12 FLUT,5.0MMX 13CM
Common Device Name
BURR, SURGICAL, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key13995224
Report Number1220246-2022-04699
Device Sequence Number17442099
Product Code GFF
UDI-Device Identifier00888867043848
UDI-Public00888867043848
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2026
Device Model NumberBUR,OVAL,FLUSHCUT,12 FLUT,5.0MMX 13CM
Device Catalogue NumberAR-8500FOT
Device Lot Number13735771
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer03/17/2022
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-