| Model Number |
Ø9X 20MM BC IF SCRW, VENTED |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament ligamentoplasty surgery the head of the device became detached from the body at the time of screwing.All broken parts were retrieved from the patient.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a different device (ar-4020c-06).It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed.Upon visual inspection it was noted that part of the device is missing.Only the head of the anchor was returned for evaluation.The cause of the event is undetermined, however likely causes include improper bone prep; misaligned insertion; prying/leveraging the device during insertion and/or shallow driver engagement depth.
|
| |
|