| Model Number |
328203 |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.A complaint lot history check was performed on lot # 1187343 for missing inner shield.This is the 1st.Related complaint for missing inner shield on lot # 1187343.A lot history review was carried out and no related non conformances were raised in association with this packaged lot concluding all inspections were performed as per the applicable operations and met qc specifications.As no samples and/or photo(s) were received the investigation concluded: unconfirmed: bd was not able to duplicate or confirm the customer¿s indicated failure as no samples or photos were returned root cause cannot be determined at this time as the issue is unconfirmed as no samples or photos were returned.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that 12 bd ultra-fine¿ original pen needles experienced product damage while still considered operable.The following information was provided by the initial reporter: consumer reported 12 missing the pink inner shields.No injury from this incident.
|
| |
|
Search Alerts/Recalls
|