• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION V-14 CONTROL WIRE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION V-14 CONTROL WIRE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 40456
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the balloon became entrapped on the guidewire.The target lesion was located in the anterior tibial artery.A 3.0mm x 150mm x 150cm sterling sl balloon catheter and a 182cm v-14 control wire were selected for use.The balloon was advanced for dilation.However, it was noted that the balloon got stuck when advancing into the guidewire.So physician decided to remove the balloon and the guidewire, but the guidewire gets trapped inside the balloon and the balloon was completely damaged when attempting to separate.The procedure was completed with another of same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that the balloon became entrapped on the guidewire.The target lesion was located in the anterior tibial artery.A 3.0mm x 150mm x 150cm sterling sl balloon catheter and a 182cm v-14 control wire were selected for use.The balloon was advanced for dilation.However, it was noted that the balloon got stuck when advancing into the guidewire.So physician decided to remove the balloon and the guidewire, but the guidewire gets trapped inside the balloon and the balloon was completely damaged when attempting to separate.The procedure was completed with another of same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: analysis of returned product revealed that the distal tip kinked.The reported issues could be confirmed.Additionally, the outer diameter was measured and the results were within specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-14 CONTROL WIRE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13996254
Report Number2134265-2022-03890
Device Sequence Number5103745
Product Code DQX
UDI-Device Identifier08714729799207
UDI-Public08714729799207
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K112745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/24/2023
Device Model Number40456
Device Catalogue Number40456
Device Lot Number0028084226
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer05/11/2022
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date06/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
Patient Weight85 KG
-
-