| Model Number |
3161 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference numbers: 1627487-2022-01837, 1627487-2022-01838, 1627487-2022-01839, 1627487-2022-01841.It was reported that the leads were found to have migrated and fractured during an inoperable ipg replacement (related manufacturer reference number 1627487-2022-00050) on (b)(6) 2022.Imaging confirmed the fracture and migration.As a result, the physician explanted and replaced all peripheral leads with two octrode leads during the same procedure.Post-operatively, stimulation therapy was restored.It is unknown which 4 leads the patient has implanted, therefore all leads are being reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A patient experiencing lead migration and fracture was reported to abbott.The patient leads were explanted and replaced.Therapy was restored.The results of the investigation are inconclusive since the devices were not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.
|
| |
|