Model Number G55814
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) #k172288.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.The device history record contains a nonconformance that could potentially be related to the cutting wire securing component separating from catheter.The device goes through various inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion omni-tome pre-loaded sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used two (2) cook fusion omni-tome pre-loaded sphincterotomes.It was initially reported that the cutting wires became "damaged/broke." there was no reportable information at this time.Per clarification received on 03/07/2022 the cutting wire anchor disconnected at the patient end of the device.[subject of report] a section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
An emdr report was initially sent on 04 april 2022, based on the information that the cutting wire detached at the patient end of the sphincterotome.The device was received for evaluation on 27 april 2022.Our evaluation of the returned device was unable to confirm the report and determined that the cutting wire securing component (anchor) moved proximal to intended location within catheter.This malfunction has not historically caused or contributed to a death or serious injury.Therefore, this no longer meets fda reporting requirements and this follow up emdr is being sent to cancel the initial emdr.
 
Event or Problem Description
This follow up emdr is being sent to cancel the initial emdr submitted.See h10 for justification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION OMNI-TOME PRE-LOADED SPHINCTEROTOME
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key13996955
Report Number1037905-2022-00153
Device Sequence Number4432454
Product Code KNS
UDI-Device Identifier00827002558144
UDI-Public(01)00827002558144(17)241222(10)W4549281
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/22/2024
Device Model NumberG55814
Device Catalogue NumberFS-OMNI-35-260
Device Lot NumberW4549281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer04/27/2022
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COOK METII-35-260 WIRE GUIDE; ERBE ELECTROSURGICAL GENERATOR; OLYMPUS DUODENOSCOPE, MODEL UNKNOWN
Patient SexUnknown
-
-