OLYMPUS MEDICAL SYSTEMS CORP. OES BRONCHOFIBERSCOPE Back to Search Results
Model Number BF-1T60
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Olympus (omsc) was informed that the customer broncho-fiberscope was returned for service/repair for a reported ¿leak¿ found during an unspecified event.However, upon inspecting and testing, the coating of the insertion tube was found peeling off.The reportable defect was found during service inspection.No death, injury or harm was reported.
 
Additional Manufacturer Narrative
The customer¿s reported issue was confirmed as the scope was leaking from a hole(s) in the insertion tube.The service evaluation also identified the coating of the insertion tube was found peeling off.Additionally, the bending angulation is below standard specification due to a stretched angle wire.The bending section if loose/wrinkled, the insertion tube is deformed/dented and has scratches, portions of the bending section cover adhesive is missing, the universal cord is deformed/dented with scratches.There are scratches on the control body, eyepiece, grip, angulation lever, up/down plate and the light guide connector.The cause of the peeling insertion tube could not be confirmed.However, the investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.If additional information becomes available a follow up medical device report will be supplemented accordingly.
 
Additional Manufacturer Narrative
Correction to date rec¿d by mfr of the initial medwatch.The aware date should be 06-mar-2022.This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that stress of repeated use, external factors, or handling cause the reported event however the root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES BRONCHOFIBERSCOPE
Common Device Name
BRONCHOFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13997539
Report Number8010047-2022-05623
Device Sequence Number4452770
Product Code EOQ
UDI-Device Identifier04953170339264
UDI-Public04953170339264
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K023984
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-1T60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date06/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-