Model Number 2090
Medical Device Problem Code Computer Operating System Problem (2898)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: analysis of the programmer was unable to confirm the customer comment that the session could not be ended.Analysis also noted that the stylus did not align with the cursor, the lower handle cover was cracked and the programmer failed left antenna test at functional testing.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the programmer had to be powered off in order to end session after the interrogation of an implantable device.The programmer was returned for service and subsequently tested out-of-specification during manufacturers analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key13997730
Report Number2182208-2022-01029
Device Sequence Number11978304
Product Code KRG
UDI-Device Identifier00613994434937
UDI-Public00613994434937
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Initial Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/23/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-