Manufacturer's investigation conclusion: the reported event of the system controller¿s power cables being damaged, resulting in an inability to communicate with the system monitor, was confirmed.The returned system controller (serial number (b)(6)) was observed to have damaged power cables upon arrival, as the cables were slightly kinked, and the inner wires were visible within the white power cable near its bend relief (controller side).The controller was opened and its power cables were replaced with test cables.After this replacement, communication with the monitor was restored.The controller was then functionally tested and was found to perform as intended when test power cables were in use.After communication with the monitor was restored, a log file was extracted from the controller containing data spanning approximately 1 day ((b)(6) 2022 ¿ (b)(6) 2022 per timestamp).A low voltage advisory alarm was observed on (b)(6) 2022 at 5:46 due to extended use of the batteries.This alarm transitioned into a hazard alarm at 5:48, then into a no external power alarm at 5:55 once both batteries had fully depleted.The pump operated on backup battery power until approximately 7:32 of the same day; at this time, the pump was observed to stop, indicating that the backup battery had been fully depleted.The pump¿s speed remained at 0 rpm throughout the remainder of the data ¿ the observed pump stop caused by full battery depletion will be addressed via the pump¿s investigation.No other notable events were observed.The system controller¿s original power cables were opened, and a fractured orange wire within the white cable was observed underneath the damaged bend relief (controller side).This wire is responsible for allowing the controller to communicate with the system monitor.The root cause of the reported event was unable to be conclusively determined through this analysis.Review of the device history record for system controller, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The heartmate 3 patient handbook, section 10 ¿safety checklists¿ instructs users to regularly inspect their equipment, including their system controllers, and to avoid using equipment that appears damaged.Users are encouraged to replace any equipment that appears damaged.The heartmate 3 patient handbook section titled "emergency contact list" cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|