• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® NO ADDITIVE (Z) TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (BROKEN BOW) BD VACUTAINER® NO ADDITIVE (Z) TUBES; SPECIMEN TRANSPORT AND STORAGE CONTAINER Back to Search Results
Catalog Number 366703
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary: bd received one (1) sample and one (1) photo for investigation.The photo was reviewed and the customer¿s indicated failure mode for foreign matter (plastic) inside of the tube was observed.Additionally, the customer sample was evaluated by visual examination and the indicated failure mode for foreign matter (plastic) inside of the tube with the incident lot was observed.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality notifications.This complaint has been confirmed for the indicated failure mode of foreign matter (plastic) inside of the tube based on the provided photo and returned sample.Bd determined that the root cause of the indicated failure mode was attributed to manufacturing.Bd has increased awareness for cleanliness and reduction of foreign matter in the manufacturing plants.Several projects are in process to help reduce the potential of foreign matter introduction into tubes.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported when using the bd vacutainer® no additive (z) tubes, the device experienced foreign matter in the tube.The following information was provided by the initial reporter.The customer stated: it was reported that tube had a plastic piece in them."plastic piece found in tube".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® NO ADDITIVE (Z) TUBES
Common Device Name
SPECIMEN TRANSPORT AND STORAGE CONTAINER
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13999479
Report Number1917413-2022-00166
Device Sequence Number5967562
Product Code FMH
UDI-Device Identifier00382903667031
UDI-Public00382903667031
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2023
Device Catalogue Number366703
Device Lot Number1259225
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Initial Report FDA Received Date04/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/16/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-