As reported by a field clinical specialist (fcs), during a right tf tavr procedure of a 23 mm sapien 3 ultra valve in the non-edwards valve, severe difficulty was noted when passing the first valve through the esheath.The valve would not advance past the femoral head.It was decided to remove the delivery system and valve from the patient.A new system and valve were opened and the valve passed through the 16f esheath without issue.Upon removal, there was a extravasation around access site observed and vascular surgery was consulted.There was no damage of the devices noted.
|
|
A supplemental mdr is being submitted for correction and additional information.The following sections of this report have been updated: corrected h.6 component codes and investigation conclusions.Added new information to h.6 type of investigation and investigation findings.The esheath plus was not returned to edwards lifesciences for evaluation.Without the device returned for evaluation, visual inspection, functional testing and dimensional analysis were unable to be completed.Imagery was provided from the site and revealed the following: the minimum luminal diameter (mld) was 4.7mm which is smaller than required for safe use of the 14f esheath (5.5mm) and no notable calcification or tortuosity present in that region.Although the work order of the complaint device was not provided, the device undergoes multiple 100% inspections during manufacturing.These inspections and tests during the manufacturing process support that it is unlikely that a non-conformance contributed to the reported complaint.The complaint was unable to be confirmed without the returned device or procedural imagery.As the device was not returned, engineering was unable to perform any visual inspection, functional testing, or dimensional analysis.Therefore, the presence of a manufacturing non-conformance was unable to be determined.A review of manufacturing mitigations supports that the sheath has proper inspections in place to detect issues related to the complaint events.A review of ifu/training materials revealed no deficiencies.Furthermore, there was no report of any issues with the sheath during device unpacking or preparation.As reported, ''as such, it is possible that patient factors (undersized vessel) likely contributed to the complaint events.However, a definite root cause was unable to be determined''.Per the procedural training manual, ''push force can vary due to angle of insertion, vessel diameter, tortuosity and degree of calcification.'' per the imaging evaluation at the femoral head, the vessel appears undersized for safe use of the 14f esheath (<5.5mm).These patient factors can create a challenging pathway for delivery system advancement through the sheath as it creates a constrained portion leading to resistance.This patient factors, in conjunction with the exposed apices of the crimped s3u thv may increase the rate of the thv getting caught on the sheath, preventing further advancement.Capa has identified to mitigate against the failure.As such, it is possible that patient factors (undersized vessel) likely contributed to the complaint events.However, a definite root cause was unable to be determined.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No ifu/labeling/training manual inadequacies were identified.A product risk assessment (pra) was previously initiated and to address the issue, a corrective action preventative action (capa) was initiated to drive corrective actions.
|