WRIGHTS LANE SYNTHES USA PRODUCTS LLC RECON LOCKING AIMING ARM FOR LATERAL ENTRY FEMORAL NAILS-EX; GUIDE, SURGICAL, INSTRUMENT
|
Back to Search Results |
|
| Model Number |
03.010.048 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Reporter is a j&j employee.The subject device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on an unknown date, during inspection of the back up devices it was discovered that the gold peg of one (1) recon locking aiming arm for lfn nails-ex has snapped off, the threads of one (1) extraction screw for ti femoral & tibial nails and one (1) extraction screw/m8 for suprapatellar are stripped, and one (1) depth gauge for 2.0/2.4 and 2.7 screws was stuck in place.There was no patient involvement.This report is for one (1) recon locking aiming arm for lateral entry femoral nails-ex.This is report 1 of 4 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: investigation summary: service and repair evaluation: the customer reported, the gold peg of one recon lckng aiming arm for lfn nails-ex snapped off.The repair technician reported, locking mechanism broken, failed cosmetics test.Damaged component is the reason for repair.The item is not repairable, per the inspection sheet.The cause of the issue is damaged component.The item will be forwarded to customer quality.The evaluation was confirmed.H4: device history: part#: 03.010.048, synthes lot#: 6074250, supplier lot#: n/a, release to warehouse date: 29 jan2009, manufactured by: synthes brandywine.No ncrs were generated, during production.Review of the device history record(s) showed that there were no issues, during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information, which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained, that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
This is report 1 of 6 for (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|