| Model Number |
1012623-39 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that an omnilink elite stent delivery system was removed from the package.It was noted that the stent had flared struts and was moved/dislocated on the delivery system balloon.The device was not used and there was no patient involvement.No additional information regarding this issue was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
Subsequent to the initial report, additional information was provided.It was reported that the absolute pro vascular stent delivery system was removed from the device packaging and prepared for use.It was noted that the back portion of the stent delivery system was split.The device was not used and there was no patient involvement.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the returned device.The reported stent dislodgement was confirmed.The reported material deformation was not confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported stent dislodgement and material deformation.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|