| Model Number |
38275 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the catheter tip detached.An imager ii angiographic 4f catheter was selected for use in a fistuloplasty.When the catheter was removed from the brachial artery, the tip detached.There was no resistance when the catheter was removed.Surgical intervention was required to open the vessel to remove the tip.The procedure had already been completed before the tip detached.The patient was stable and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
G1: mfr site state: (b)(6).Device eval by mfr: the returned product consisted of an imager ii catheter.25mm of tip was separated from the shaft.The separated ends appeared to be jagged as if the separation was do to tensile/ductile forces were applied.There were marks on the separated section which appeared to look like forceps imprints.A x-ray of the device was completed to see if there were any braids that were stretched.The reported tip separation was confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that the catheter tip detached.An imager ii angiographic 4f catheter was selected for use in a fistuloplasty.When the catheter was removed from the brachial artery, the tip detached.There was no resistance when the catheter was removed.Surgical intervention was required to open the vessel to remove the tip.The procedure had already been completed before the tip detached.The patient was stable and no further complications were reported.
|
| |
|
Search Alerts/Recalls
|