| Model Number |
AAB00 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Date of event: date unknown/ not provided.Implant date: lens was not implanted.Explant date: lens was not implanted, therefor not explanted.(b)(6).The intraocular lens (iol) has not yet been returned for evaluation.Therefore, a failure analysis of the complaint device could not be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review and possible product return and evaluation, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information.However; to date, no response has been received.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the internal packaging containing the lens was found opened and non-sterile.Through follow up we learnt that the issue was not due to shipping damage or user error.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9 - device available for evaluation? yes.Section d9 - date returned to manufacturer: 9th may 2022.Section h3 - device evaluated by manufacturer? yes.Device evaluation: the complaint product was received in its original folding carton package and there was no damage observed in the box.The intraocular lens with the case was inside an opened inner pouch.The pouch was visually inspected under fluorescent magnifier lamp by a certified operator at the first pack area.It was observed seal discontinuity on the mylar film side indicating lack of cohesion which could have caused weak seal condition.A failure investigation (fi) was conducted to further investigate the complaint issue.Device history record review was conducted results were within specification and no abnormal situation was identified.A search of complaints related to the production order was performed on june 10, 2022.The search revealed that no other complaint was received from this production order.Calibration record and preventive maintenance record for the urania sealer used to perform the sealing process was reviewed, and there was no abnormal situation identified during the calibration and preventive maintenances processes.Manufacturing controls are in place and require all pouches of the production order must be sealed and inspected under the fluorescent magnifier, using the magnification and illumination features to assure seal is acceptable as this guarantees the sterility barrier of the units.Additionally, the sealer parameter¿s verification must be performed prior to starting the sealing process at the beginning of each shift.A sealer periodic iol creep & burst test is performed every 2 hours + 20 minutes or as required.If any of the pouches fail to meet specifications, the area supervisor or quality specialist must be notified.Based on the analysis of the returned device, the complaint issue reported was considered verified and determined to be an isolated event.Awareness communication was provided to employees of first pack and boxing area to reinforce the procedures requirements.Also retraining was given to the operators involved.These types of events will continue to be monitored.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|