• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, MULTIFOCAL INTRAOCULAR Back to Search Results
Model Number DFW150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Headache (1880); Blurred Vision (2137); Visual Disturbances (2140)
Date of Event 09/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity: asku.The device was not returned for evaluation.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the preloaded intraocular lens (iol) was explanted from patient's left eye due to headaches, light sensitivity and blurry vision.Another competitor lens was used as the replacement iol.There was no patient injury and no other intervention was required.The patient was doing good during post-operation day 1 visit.No further information was provided.
 
Additional Manufacturer Narrative
Additional info: section d9: device available for evaluation: yes section d9: returned to manufacturer on: mar 25, 2022 section h3: device evaluated by manufacturer: yes device evaluation: the complaint lens was received stuck to a piece of styrofoam.Visual inspection under magnification revealed that the lens was cut in half.The lens was cleaned, and no further issues were observed.Due to the returned condition of the lens (cut in half) no further product evaluation could be performed.Conclusion: as a result of the investigation there is no indication of a product quality deficiency, and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14004605
Report Number3012236936-2022-00833
Device Sequence Number4372506
Product Code MFK
UDI-Device Identifier05050474751156
UDI-Public(01)05050474751156(17)240309
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/09/2024
Device Model NumberDFW150
Device Catalogue NumberDFW150U145
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer05/12/2022
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date05/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexFemale
Patient Weight77 KG
Patient RaceWhite
-
-