| Model Number |
24017545 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Vertebral Fracture (4520)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed by review of the radiographic image provided.No product was returned to nuvasive for evaluation as the device remains in-situ.Operative notes were not provided for review of usage/technique.A definitive root cause was unable to be determined with the information provided.There are warnings in the device labeling that this type of event can occur: "potential adverse events and complications: potential risks identified with the use of this system, which may require additional surgery, include: fracture of the vertebra"."warnings, cautions and precautions: the implantation of spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient".If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial posterior fixation procedure on an unknown date.Subsequently, post-operative imaging obtained on an unknown date identified fracture of the patient's right pedicle at l4.The surgeon believes the fracture occurred during insertion of the screw.There is no intent to revise the implanted construct at this time and the patient has no adverse symptoms.No further patient impact was reported.
|
| |
|
Search Alerts/Recalls
|