• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. PROCISE LW COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. PROCISE LW COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number EIC7070-01
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pneumonia (2011); Respiratory Arrest (4461); Wheezing (4463); Upper Respiratory Tract Injury (4599)
Date of Event 01/01/2021
Type of Reportable Event Death
Event or Problem Description
It was reported that on literature review the management of infantile tongue base cyst with laryngomalacia, 5 patients had complications after a laryngomalacia procedure using a procise lw coblation wand.1 patient had a disconnection of the tube on the 1st day after the operation, and was not reinserted because of stable breathing and disappearance of stridor at that time, laryngeal stridor appeared again on the 5th day after the operation, it required a revision surgery 1 month after the operation and was intubated for 5 days and recovered well, 3 patients had an intubation time extended to 8 days due to pneumonia, 1 other patient was extubated postoperatively, and on the first postoperative day, laryngeal stridor reappeared, and laryngoscopy revealed epiglottis inversion.The patient was intubated in the icu.The tube was disconnected on the 4th day and the patient died.The reason may be that the tracheal tube was mistakenly inserted into the surgical area of the lingual root cyst during emergency intubation due to the edema and unclear structure of the supraglottic tissue after coblation.This resulted in subcutaneous emphysema in the neck caused by the air flow through the anterior epiglottic space during balloon pressurization for oxygen administration, which aggravated the asphyxia and led to the death of the child.Patients outcome are unknown, except for the one that died.No further information is available.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).Paper: the management of infantile tongue base cyst with laryngomalacia 10.13201/j.Issn.2096-7993 2021 06 007.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states the images provided in the article have been interpreted within the text; therefore, no further analysis of the images is required, and it cannot be concluded that there was a malperformance of the s&n device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Internal complaint reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCISE LW COBLATOR II
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14004862
Report Number3006524618-2022-00190
Device Sequence Number4339974
Product Code GEI
UDI-Device Identifier00817470000443
UDI-Public00817470000443
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K202006
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Consumer,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEIC7070-01
Device Catalogue NumberEIC7070-01
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer05/15/2023
Initial Report FDA Received Date04/04/2022
Supplement Report FDA Received Date05/16/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Other;
Patient SexUnknown
-
-