Model Number MMT-1715KL
Device Problem Power Problem (3010)
Patient Problems Hypoglycemia (1912); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/02/2022
Event Type  malfunction  
Event Description
Information received by medtronic indicated that the insulin pump alarmed critical pump error and displayed open book image on the insulin pump screen.No harm requiring medical intervention was reported.Troubleshooting was not performed, the customer will discontinue the use of device.
 
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Manufacturer Narrative
The insulin pump was received with a critical pump error (open book image) alarm.Unable to perform the displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement test, active current measurement test and self test due to critical pump error (open book image) alarm.Unable to verify low bgs.The insulin pump was cut open to perform visual inspection and found moisture damage on electrical board 1 and the force sensor.Successfully utilized crest and thus to download history files, traces and comlink3 files.Critical pump error (open book image) alarm confirmed and was triggered by a pump error 35 fatal alarm confirmed in the history files on (b)(6)2022 7:15:00 am due to moisture damage on the force sensor.Test p-cap and reservoir locked properly into reservoir compartment during testing.The following were noted during visual inspection: pillowing keypad overlay, cracked keypad overlay, stained keypad overlay, keypad overlay texture damage, scratched case, missing display window cover, cracked case - corner of belt clip rails, cracked case (battery tube), battery tube threads - cracked and corroded battery tube.Unable to verify low bgs.Critical pump error (open book image) alarm confirmed due to pump error 35 fatal alarm.Pump error 35 fatal alarm confirmed due to moisture damage on the force sensor.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
630G INSULIN PUMP MMT-1715KL 630G
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key14009411
MDR Text Key290295063
Report Number2032227-2022-175680
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000763000316655
UDI-Public(01)000000763000316655(17)230604
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/04/2023
Device Model NumberMMT-1715KL
Device Catalogue NumberMMT-1715KL
Device Lot NumberHG4CUCJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/02/2022
Initial Date FDA Received04/05/2022
Supplement Dates Manufacturer Received09/09/2022
Supplement Dates FDA Received09/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/04/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age36 YR
Patient SexFemale
Patient Weight100 KG
-
-