• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT CAPTIVIA - CW; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT CAPTIVIA - CW; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VAMF4242C200TU
Medical Device Problem Codes Defective Component (2292); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
A valiant captivia stent graft was intended to be implanted in the endovascular treatment of a thoracic aortic aneurysm.It was reported that during the index procedure, a gap between the distal edge of the graft cover and the tapered tip was noticed and the procedure was aborted.It was confirmed the stent graft was inspected prior to use and the gap was noted pre its insertion into the patient.Per physician, the cause of the event was due to the patient's anatomy and tight calcified iliac access.It was also noted that the gap may have contributed to device not being able to pass through the common iliac artery to its intended implant site in the thoracic aorta but the gap was present outside the patient prior to insertion.No additional clinical sequelae were reported and the patient will be monitored.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - CW
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14011470
Report Number9612164-2022-01318
Device Sequence Number14781731
Product Code MIH
UDI-Device Identifier00643169346536
UDI-Public00643169346536
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/28/2023
Device Model NumberVAMF4242C200TU
Device Catalogue NumberVAMF4242C200TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Initial Report FDA Received Date04/05/2022
Date Device Manufactured05/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age70 YR
Patient SexFemale
-
-