| Catalog Number |
03P89-24 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2022 abbott point of care was contacted by a customer regarding i-stat pt/inr cartridges that yielded suspected discrepant pt/inr results of 8.0 on a 53 year old female patient.There was no additional patient information available at the time of this report.Return product is not available for investigation.Method: i-stat; lab.Date: 22-mar-2022 ; 22-mar-2022.Pt/inr: 8.0; 4.7.Sample : venous; venous.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: # (b)(4).The investigation was completed on (b)(6).2022.A review of the device history record (dhr) confirmed the cartridge lot passed release specifications.Pt/inr cartridge lot s21269c expired on 25-mar-2022 prior to elevation of incident to level 2 investigation.No deficiency has been identified.
|
| |
|
Event or Problem Description
|
|
Na.
|
| |
|