• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DALE MEDICAL PRODUCTS, INC. DALE TRACHEOSTOMY TUBE HOLDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DALE MEDICAL PRODUCTS, INC. DALE TRACHEOSTOMY TUBE HOLDER Back to Search Results
Model Number 240
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2021
Type of Reportable Event Malfunction
Event or Problem Description
"adolescent patient with tracheostomy tube and mechanical ventilation.While doing trach care, velcro part of the trach.Holder / strap broke off completely from the soft part of the strap that encircles the neck.Staff member changed to a new holder and ensured it was secure.Thirty minutes later, the velcro part broke again, and holder was replaced again.No actual patient harm but potential for trach tube dislodgement.".
 
Additional Manufacturer Narrative
It was reported that on (b)(6) 2021 there was an adolescent patient with a tracheostomy tube and mechanical ventilation, and while doing trach care, the velcro part of the trach holder / strap broke off completely from the soft part of the strap that circles the neck.A staff member changed to a new holder and ensured it was secure.Thirty minutes later, the velcro part broke again, and the holder was replaced again.No patient harm was reported.On the same day, we received a second report of another issue of an adolescent patient with a tracheostomy tube and mechanical ventilation, but on 12/26/2021.We contacted the initial reporter to get additional information about both events and the devices: the initial reporter confirmed that both reports were relative to the same patient.The initial reporter confirmed that the facility did not retain any of the broken devices for manufacturer review and the lot numbers for the (b)(6) 2021.Event were unknown, but the lot number for the (b)(6) 2021.Event is 21h2318.The initial reporter could not confirm if the (velcro) tabs had broken or if they had come unsewn from the neck strap.The initial reporter confirmed that she did check with other departments at johns hopkins that uses the dale 240 device to see if there were any other issues noted and there were not.Without having defective or suspect devices to review or even the actual lot number associated with this event, dale medical cannot confirm a cause for the issue.No additional information is available.Due to the reported incident and in an abundance of caution, this medwatch is being filed.If any further relevant information is identified or obtained, a supplemental medwatch will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DALE TRACHEOSTOMY TUBE HOLDER
Common Device Name
TRACHEOSTOMY TUBE HOLDER
Manufacturer (Section D)
DALE MEDICAL PRODUCTS, INC.
40 kenwood circle
suite 7
franklin MA 02038 3298
Manufacturer (Section G)
DALE MEDICAL PRODUCTS, INC.
40 kenwood circle
suite 7
franklin MA 02038 3298
Manufacturer Contact
raymond gruneberg
40 kenwood circle
suite 7
franklin, MA 02038-3298
5083164608
MDR Report Key14013346
Report Number1214422-2022-00001
Device Sequence Number11915402
Product Code CBH
UDI-Device Identifier00817612020667
UDI-Public00817612020667
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
LISTING ONLY
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 01/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number240
Device Catalogue Number240
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age14 YR
Patient SexMale
Patient Weight21 KG
-
-