ZIMMER BIOMET, INC. G7 NEUTRAL E1 LINER 32MM F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Unspecified Tissue Injury (4559)
Date of Event 01/11/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a total hip study patient (36) underwent revision approximately 3 years post implantation due to painful psoas impingement confirmed by ultrasound.No additional information.
 
Additional Manufacturer Narrative
(b)(4).Source: foreign: country: (b)(6).Reported event was unable to be confirmed due to limited information received from the customer.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.Doi: does the use of the largest possible metal head increase the wear of vitamin e-doped cross-linked polyethylene? two-year results from a randomized controlled trial.10.1302/0301-620x.103b7.Bjj-2020-2064.R1.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2022-00264 and 0001825034-2022-00800.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 32MM F
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14013571
Report Number0001825034-2022-00799
Device Sequence Number12034710
Product Code PBI
UDI-Device Identifier00880304526365
UDI-Public(01)00880304526365(17)201201(10)3697412
Combination Product (Y/N)N
Initial Reporter CountryDA
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Literature,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/01/2020
Device Model NumberN/A
Device Catalogue Number010000850
Device Lot Number3697412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CAT# 010000664 LOT# 3659134 G7 SHELL; CAT# 163668 LOT# 00J3698303 HEAD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-